::: UniChrom BEL RUS DEU ENG
Main Page / Products / < Back 

High performance liquid microcolumn chromatographic system "Milichrom-6" 5,4

Producer: SDB "Nauchpribor" (Scientific Instruments)

Microcolumn - very little HPLC column with small eluent expends. Liquid
Chromatograph. Mi+Li+Chrom = Milichrom

UniChrom automatically detects 1(2) pump system. It is possible acquiring at several wavelengths simultaneously in range from 190 to 360 nm. When gradient program is stopped is possible the scanning of absorption (fluorescence) spectrum of substance currently in photometric cell.

Instrument installation manual

Microcolumn liquid chromatographs MiliChrom-6, -5, -4 and UniChrom - RUS (PDF)
the manual currently in Russian only, so if you are interested in getting English version - do not hesitate - contact us

Highlights

  • Monoblock
  • Spectral range from 190 to 360 nm
  • Smallest microcolumn HPLC system in the world
  • Range of eluent flow for chromatographic mode - 0.002 to 0.5 ml/min
  • Very low eluent consumption
  • Multiwavelenght spectrophotometer with photomultiplier tube
  • Chromatogram can be acquired at 5 wavelnghts simultaneously.
  • Two or Single pump configuration
  • Autosampler for 29 vials
  • Low price complete HPLC system

System work style examples

Polycyclic compounds on Milichrom-5 Metall salts on photometric detector Metall salts - spectral reference Multiwave HPLC of different drinks

Specifications

Detector: two-beam multiwavelength spectrophotometer.
Wavelenght range 190-360 nm
Simulatneous detection wavelenghts number Up to 8
Wavelength settings error 0.5 nm
Basleline noise level 1E-4 o.d.u.
Baseline drift level 5E-5 o.d.u./hour
Limit of acceptable value of mean squared
deviation of output signal level
1%
Pump: syringe; dosing; syringe volume: 2500 mkl
Eluent flow speed 5-999 mkl/min
Relative error of eluent flow speed not more 1%
Maximal pressure 7 MPa
Eluation mode Stepped gradient, Gradient
Columns 2x64 -120 mm; silasorb C18; silasorb 600
Autosampler automated, programmable, for 29 samples
Sample vial volume 200 mkl
Injection volume 1-99 mkl stepped through 1 mkl
Corrosion protection stainless steel 12Cr18Ni10Ti; glass; quartz; teflon-4
Size and weight: 330x450x710 mm; weight 17 kg

Networked Chromatograph This instrument driver is designed as "Networked Chromatograph"